Home

Peptide Testing Guides Knowledge Center

How peptides are tested, what each result means and where its limits are. Short reads by the people who run our lab.

Our autosampler picks up a sample vial from the tray and carries it to the injection port
INSIDE THE LAB · FROM TRAY TO INJECTION · 0:38

BROWSE BY CATEGORY

Lab terms

The words you will see on a report, in one or two sentences each.

High-performance liquid chromatography. It separates the compounds in a sample, and each one shows up as a peak.
The graph an HPLC run produces: detector signal over time. Each peak is a compound.
When a compound comes off the column. With the same method, the same compound comes off at the same time.
The share of the UV signal that belongs to the main peak. It is not the amount of peptide in the vial.
Liquid chromatography with a mass spectrometer as detector. It measures the mass of what comes off the column.
Whether the main peak is the compound on the label. Confirmed only when retention time and mass both match the reference standard.
How much peptide is in the vial, in mg, measured against a reference standard.
The smallest impurity listed separately on a report: 0.10 % of the UV signal.
READY WHEN YOU ARE

Test your own sample

Purity from $129. Identity and content by LC–MS from $179.

INSIDE THE LAB · FROM TRAY TO INJECTION · 0:38